| Technical data |
MP100H (voltage outputs) | MP400H (current outputs) |
| Configurations |
- Separate temperature probe with active linear output signal (standard)
- Separate temperature probe with direct 2-, 3- or 4-wire RTD connection (optional)
- No separate temperature probe - temperature output provided by RTD from HygroClip (optional)
|
| Humidity Temperature Plug-in Probe |
HygroClip S3 (included with MPH) |
| Humidity sensor: ROTRONIC Hygromer® C94 (capacitive) |
Temperature sensor: RTD Pt100 Ohm, 1/3 DIN |
Sensor protection: Wire mesh filter (standard) Foam filter (optional) |
| Separate Temperature Probe |
Stainless steel rod 118 x 3 mm (4.64 x 0.12") |
Pt100 1/3 DIN RTD (standard) Pt100 1/5 DIN RTD (optional) Pt100 1/10 DIN RTD (optional) |
| Operating/Measuring range |
0...100%RH -40...60°C / -40...140°F |
| Accuracy at 23°C / 73°F |
<±1.5%RH and ±0.2°C (valid for active outputs) |
| Repeatability |
better than 0.5%RH and 0.1°C |
Measuring ranges
(active outputs) | 0...1V = 0...100%RH | 4...20mA = 0...100%RH |
0...1V = -40...60°C (standard) 0...1V = -30...70°C (optional) -0.4...0.6V = -40...60°C (optional) | 4...20mA = -40...60°C (standard)
4...20mA = -30...70°C (optional) |
| Active Outputs Load |
> 1000 Ohm | < 500 Ohm |
| Supply Voltage |
3.6...35VDC, 10 mA | 5VDC + 0.02 x load, 45 mA max. |
| Cable Length Compensation |
effective for up to 100m/300ft | N/A (current outputs) |
| Connecting Cable |
standard: 3m (9ft) PUR cable with tinned ends |
| Weight |
210 g / 0.47 lb |